UV cutoffs for common solvents
The table below shows the UV cutoff (the wavelength at which the absorbance
of the solvent is equal to 1 AU) for some common chromatographic solvents.
Operating at a wavelength near or below the cutoff increases baseline noise
because of the absorbance of the solvent.
UV cutoff wavelengths for common chromatographic solvents
Solvent
1-Nitropropane
2-Butoxyethanol
Acetone
Acetonitrile
Amyl alcohol
Amyl chloride
Benzene
Carbon disulfide
Carbon tetrachloride 265
Chloroform
Cyclohexane
Cyclopentane
Diethyl amine
Dioxane
Ethanol
Ethyl acetate
Ethyl ether
Ethyl sulfide
Ethylene dichloride
UV cutoff
Solvent
(nm)
380
Ethylene glycol
220
Iso-octane
330
Isopropanol
190
Isopropyl chloride
210
Isopropyl ether
225
Methanol
280
Methyl acetate
380
Methyl ethyl ketone
Methyl isobutyl
ketone
245
Methylene chloride
200
n-Pentane
200
n-Propanol
275
n-Propyl chloride
215
Nitromethane
210
Petroleum ether
256
Pyridine
220
Tetrahydrofuran
290
Toluene
230
Xylene
UV cutoff
(nm)
210
215
205
225
220
205
260
330
334
233
190
210
225
380
210
330
230
285
290
Solvent considerations
103