4.
Remove the leak sensor from its reservoir by grasping it by its serrations
and pulling upward on it.
If you cannot easily manipulate the leak sensor after removing it
Tip:
from its reservoir, detach the leak sensor connector from the front of the
instrument (see
5.
Use a nonabrasive, lint-free wipe to dry the leak sensor prism.
40
To avoid damaging the leak sensor, do not grasp it by the
Caution:
ribbon cable.
"Replacing the detector's leak sensor" on page
TP02891
Serrations
Prism
Lint-free wipe
43).