Monitor And Lcd Adjustments; Crt Monitor Contrast And Brightness Adjustment; Pcvic/Dgvic Part Number Check; To Adjust The Contrast And Brightness - GE LOGIQ 7 Service Manual

Hide thumbs Also See for LOGIQ 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
2286865, R
IRECTION
EVISION
Section 6-5

Monitor and LCD Adjustments

6-5-1

CRT Monitor Contrast and Brightness Adjustment

Before calibrate the Monitor, check your PCVIC or DGVIC Part Number on the Utility Screen. The typical
setting value changes with its Part Number.
(To enter this screen, select Utility > System > About > Additional About Information.)

To adjust the contrast and brightness:

1.) Select Calibration from the Test Pattern Utility Touch Panel. The test pattern consists of a small
6-10
14
Figure 6-10 PCVIC/DGVIC Part Number Check
box inside a larger box.
Figure 6-11 Calibration
Section 6-5 - Monitor and LCD Adjustments
LOGIQ™ 7 S
M
ERVICE
ANUAL