CRT Monitor Contrast and Brightness Adjustment
2.) Press the Toggle button (1) for contrast and brightness. Confirm that the contrast (or
3.) Set the Contrast according to the table below.
4.) Set the Brightness at 100. Then slowly decrease the Brightness until the inner box is no longer
Table 6-3 Contrast Recommended Setting
Room Condition
Dark room for Radiology/
Dim room for Radiology/
Bright room for OB
Dark room for Cardiology
Record the final brightness and contrast settings and leave this information withe the system. Generally
speaking, do not change the controls once they have been set, the display becomes the reference for
the hard copy device(s).
NOTE:
After readjusting the monitor's Contrast and Brightness, readjust all preset and peripheral
settings.
14
brightness) indicator is displayed on the monitor. If the brightness is displayed, press the toggle
button again.
Figure 6-12 Monitor Adjustment buttons
visible. Please note that the monitor response to this adjustment logs behind the button push.
for PCVIC other than
Cardiology
Cardiology
Section 6-5 - Monitor and LCD Adjustments
(cont'd)
2
1
3
Monitor Adjustment
Contrast
2349225
for PCVIC 2349225
50
60
70
60
LOGIQ™ 7 S
Contrast
Contrast for DGVIC
2349225-2 or later
45
55
65
55
M
ERVICE
ANUAL
50
60
70
60
6-11