Probes Tested For Isolation (Sink) Current - GE LOGIQ 7 Service Manual

Hide thumbs Also See for LOGIQ 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
2286865, R
IRECTION
EVISION
10-7-9-5
Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont'd)
8.) Test every probe and record in Table 10-26 the test results for every probe.
The test passes when all readings measure less than the values in Table 10-30.
9.) Keep a record of the results with other hard copies of Planned Maintenance data.
.
Table 10-26 Probes Tested for Isolation (Sink) Current
14
Probe Tested
Section 10-7 - Electrical Safety Tests
LOGIQ™ 7 S
ISO/Mains Applied
M
ERVICE
ANUAL
10-29