GE H
EALTHCARE
D
2286865, R
IRECTION
EVISION
6-5-5-2
For Technart (cont'd)
4.) Htotal Check:
a.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
b.) Check HTotal value:
-
-
c.) Repeat to press the MENU button until the "HTotal" blinks.
d.) Repeat to press the PLUS or MINUS button to set 16.
e.) Press the AUTO button once to save the setting.
f.) Repeat to press the MENU button until the "Exit" blinks
g.) Press the AUTO button once to exit from the menu.

5.) Hposition/VPosition Check:

a.) Close the paint software and display windows desktop.
6-24
14
If it is 16, go to step5 - Hposition/VPosition check
If it is NOT 16, go to next step.
Figure 6-31 H Total
Figure 6-32 Start-up Screen
Section 6-5 - Monitor and LCD Adjustments