Reloading the Probe Data

This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details,
refer to Section 7, Troubleshooting.
This rewrites the probe data (probe delay data for beam forming) saved on the flash memories of DDBF
and TRAP boards.
1.) Insert the Service Dongle into the USB port, located at the rear panel of the scanner.
Service Dongle
2.) Power ON the scanner.
3.) Refer to
4.) Windows Desktop screen is displayed.
5.) Erase the file, "D:\Pegasus\target\resources\Pegasus\ProbeInfo.res."
6.) Erase the above file completely from Recycle bin, then restart the scanner.
14
Figure 6-9 Inserting the Service Dongle
4-2-2-3 Entering Maintenance Mode