Gems Leakage Current Limits; Chassis Leakage Current Limits—Accessible Metal Surfaces - GE LOGIQ 7 Service Manual

Hide thumbs Also See for LOGIQ 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
2286865, R
IRECTION
EVISION
10-7-2

GEMS Leakage Current Limits

The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment
Safety Standards. These limits are GEMS standards and in some cases are lower than the above
standards listed.
Table 10-14 Chassis Leakage Current Limits—Accessible Metal Surfaces
Country
USA
Other
Table 10-15 Type BF Applied Part Leakage Current Limits - Non-Conductive
(Floating) Surface and Cavity Probes
Country
USA
Other
Table 10-16 Type CF Applied Part Leakage Current Limits - Surgical Probes and
ECG Connections
Country
USA
Other
NOTE:
*Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the
part to determine the amount of current that will pass (or sink) to ground if a patient contacted
mains voltage.
The following tests are performed at the factory and should be performed at the site. These tests are:
grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All
measurements are made with an electrical safety analyzer Model 600/600E built by Dale Technology
Corporation or equivalent device.
14
Normal Condition
N/A
0.1 mA
Normal
Condition
Open Ground
0.05 mA
0.05 mA
0.1 mA
Normal
Condition
Open Ground
0.01 mA
0.05mA
0.01 mA
0.05 mA
Section 10-7 - Electrical Safety Tests
Open Ground
Reverse Polarity
0.3 mA
0.5 mA
Reverse Polarity
0.05 mA
0.5 mA
0.5 mA
Reverse Polarity
0.05 mA
0.05 mA
LOGIQ™ 7 S
ERVICE
Open Neutral
0.3 mA
N/A
0.5 mA
0.5 mA
Open Neutral
*Mains Applied
0.05 mA
N/A
0.5 mA
5.0 mA
Open Neutral
*Mains Applied
N/A
0.025 mA
0.05 mA
0.05 mA
M
ANUAL
10-13