3.5.1 Preliminary checks – MONITORING PARAMETERS
Falco 202 Evo
Based on the PRP set by the User and on the patient's [patient simulator]
characteristics, the lung ventilator is able to monitor and measure a
series of values necessary for the patient's clinical evaluation.
Before checking the value of the set parameters, leave the lung ventilator
on for at least 15 minutes. This way the system will be able to reach its
operating condition.
5. Check the compliance
between the parameters set
and those monitored.
In the middle of the screen the
system displays the operating
curves.
On the right side of the screen
you can see the monitored
parameters.
6. Change the PRP values.
PEEP: 5, 10 cmH2O
Tr. I: -2 cmH2O, 3 L/min
O2: 60%
7. Check the correspondence
between the monitored
parameters and the displayed
curves.
8. Select the key to display all the
parameters.
3-29