Performing User Calibration

Using the CX1101A/CX1102A/CX1103A
Using the CX1104A
Using the CX1105A
Using the CX1151A
User calibration allows you to perform error correction of measurement data.
This is effective for making more accurate measurements. This chapter describes
how to perform the user calibration.
The user calibration must be performed after warming-up of 30 minutes.
The user calibration is effective for:
the environmental temperature of 0 to 40 C
3 C from the temperature when the last user calibration is performed
24 hours since the last user calibration is performed
All of the user calibration data is cleared by applying the factory default. The
data is not cleared by turning the CX3300 off.