Verifying Basic Operation; If You Use The Cx1101A/Cx1102A/Cx1103A; Required Accessories; Procedure - Keysight CX3300 Series User Manual

Device current waveform analyzer
Table of Contents

Advertisement

Verifying Basic Operation

The following procedures verify the basic operation which monitors the
Calibration Output signal.
If the CX1100 has not been connected yet, connect it. See
CX1100" on page

If You Use the CX1101A/CX1102A/CX1103A

Required accessories

Procedure

1. Connect the current sensor to the Aux Out terminal on the CX3300 front
2. Press the
3. Press the
4. Set the following conditions on the Configuration dialog box (Configuration >
Keysight CX3300 User's Guide, Edition 4
"If You Use the CX1101A/CX1102A/CX1103A"
"If You Use the CX1104A"
"If You Use the CX1105A"
21.
CX1203A sensor head, 1 ea., furnished with the CX1101A/CX1102A, no need
for the CX1103A
Adapter (SMA plug to BNC jack), 1 ea., furnished with the current sensor
Coaxial cable (BNC plug to BNC plug), 1 ea., furnished with the CX3300
panel by using the accessories listed above.
(yellow),
1
channel connected to the Aux Out terminal and enable the channel.
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
Calibration Output).
Output State: ON
Source Shape: DC
Source Mode: Current
Current: 2 mA
(green),
(blue), or
2
3
Setting Up the CX3300

Verifying Basic Operation

"Connecting the
(red) key associated with the
4
23