Using the CX1104A
The CX3300 provides the offset user calibration for the CX1104A.
The user calibration data, Offset [A], is acquired and is used for error correction.
Effective for reducing the offset error.
Preparation:
Before starting the user calibration, connect the CX1104A to the CX3300 analog
input channel.
Prepare a resistive sensor head (CX1211A, CX1212A, CX1213A, CX1214A,
CX1215A, or CX1216A).
After the connection, warm up them for 30 minutes.
Proced ure:
You can perform the user calibration as shown below.
1. Press the
2. Click User Calibration to display the Configuration > User Calibration screen.
3. Click the Execute button for the channel under user calibration.
Keysight CX3300 User's Guide, Edition 4
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
You will see a dialog box which asks you to connect a resistive sensor head to
the CX1104A input using the extension cable. Then connect them as shown
in
Figure 3-5
and click Yes.
The user calibration will start.
Performing User Calibration
Using the CX1104A
95