Using The Cx1105A - Keysight CX3300 Series User Manual

Device current waveform analyzer
Table of Contents

Advertisement

Using the CX1105A

The CX3300 provides the offset user calibration for the CX1105A.
The user calibration data, Offset [V], is acquired and is used for error correction.
Effective for reducing the offset error.
Preparation:
Before starting the user calibration, connect the CX1105A to the CX3300 analog
input channel.
Prepare the test adapter.
After the connection, warm up them for 30 minutes.
Proced ure:
You can perform the user calibration as shown below.
1. Press the
2. Click User Calibration to display the Configuration > User Calibration screen.
3. Click the Execute button for the channel under user calibration.
Keysight CX3300 User's Guide, Edition 4
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
You will see a dialog box which asks you to connect the test adapter to the
CX1105A input. Then connect them as shown in
The user calibration will start.
Performing User Calibration
Using the CX1105A
Figure 3-6
and click Yes.
97