8. Perform a negative current measurement for both High Resolution mode ON
Figure 3-4
Keysight CX3300 User's Guide, Edition 4
and OFF as shown below.
a. Press the
key located in the Trigger control on the CX3300 front
Menu
panel. And set the following values on the Setting > Trigger screen.
•
Slope: Negative
Level: 5 A
•
b. Control the current source (Keysight B2900) and apply the current of the
following condition.
Condition:
•
Pulse base: 0 A
Pulse peak: 10 A
•
•
Pulse width: 1 ms
Duty cycle: 2.5% for the Keysight B2900 series
•
c. Control the CX3300 and acquire the waveform as same as the step 3-c.
And read the marker Y value then record it as follows.
•
Ion_negative=Y
Negative Pulse Acquisition
Using the CX1101A/CX1102A/CX1103A
X1 line
Y
Performing User Calibration
X2 line
93