Setting Up the CX3300
Verifying Basic Operation

Procedure

1. Set the slide switch on the test adapter to the 50 OHM. See
2. Connect the coaxial cable between the BNC jack connector on the test
3. Connect the CX1105A to the connection posts on the test adapter as shown
4. Press the
5. Press the
6. Set the following conditions on the dialog box (Configuration > Calibration
7. Click the close (x) button located in the upper right corner on the
8. Press the
9. Use the Vertical control knob to adjust the vertical scale. You will see square
Figure 1-6
26
adapter and the Aux Out terminal on the CX3300 front panel.
in
Figure
1-6.
(yellow),
1
channel connected to the Aux Out terminal and enable the channel.
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
Output).
Output State: ON
Source Shape: Square
The +5 V square pulse (1 kHz, 50% duty cycle) is applied from the Aux Out
terminal which has 50  output impedance.
Configuration dialog box to close the dialog box.
key to start waveform acquisition.
Run
pulse (approximately +2.5 V, divided by two 50 ) as shown in
Test Adapter and Calibration Output
50 OHM
Slide switch
Connection posts
+
for connecting
the CX1105A
GND
-
BNC jack connector
(green),
(blue), or
2
3
Keysight CX3300 User's Guide, Edition 4
Figure
(red) key associated with the
4
1-6.
Figure
1-6.