Setting Up the CX3300
Maintenance
Figure 1-9
34
CX1152A Troubleshooting
Replace the grabber
of the failing channel with
a "working" grabber.
No
Replace the probe lead
of the failing channel with
a "working" probe lead.
No
a "working" flat cable
Replace the flat cable
with the "working" flat cable.
START
Does the problem
still exist?
Yes
Does the problem
still exist?
Yes
Do you have
No
No
for CX1152A?
Yes
Does the problem
Yes
No
still exist?
No
END
Keysight CX3300 User's Guide, Edition 4
Flat cable and/or pod
is defective.
Contact Keysight sales office
for further troubleshooting.
Pod is defective.
Contact Keysight sales office.