Front Panel Overview - Keysight CX3300 Series User Manual

Device current waveform analyzer
Table of Contents

Advertisement

Front Panel Overview

This section describes the front panel of the CX3300.
Figure 2-1
1. Standby switch
2. Aux Out terminal
Keysight CX3300 User's Guide, Edition 4
Front Panel
1
2
Turns the instrument on. Pressing the button in the ON state makes it in the
standby state. The green LED lights when it is in the ON state.
BNC connector used to apply one of the following reference signals. The
output signal can be set by using the Configuration dialog box. For the square
pulse, the base voltage is 0 V and the output impedance is 50 .
DC current: -200 mA to +200 mA
DC voltage: -7 V to + 7 V
Square: +5 V square pulse, 1 kHz, 50% duty cycle
Pulse: +5 V square pulse, 20 ns to 1 ms period, programmable duty cycle
14
3
Using the CX3300

Front Panel Overview

10
9
12
11
13
8
7
6
5
4
39