Keysight CX3300 Series User Manual page 89

Device current waveform analyzer
Table of Contents

Advertisement

Proced ure:
1. Turn the CX3300 and the current source (Keysight B2900) on.
2. Attach the CX1206A to the CX1101A.
3. Connect the CX1101A to the CX3300.
4. Warm up the equipments. Required warm-up time is shown below.
5. Connect the equipments as shown below. Also see
6. Disable the user calibration for the channel connected to the
7. Perform a positive current measurement for both High Resolution mode OFF
Keysight CX3300 User's Guide, Edition 4
CX3300: 30 minutes after connecting the CX1101A
B2900: 60 minutes after power-on
a. Connect a BNC-banana adapter to the B2900 Force terminals. Then the
ground marker must come to the Low side.
b. Connect a BNC-banana adapter to the CX1206A input. Then the ground
marker must come to the  (minus) side.
c. Connect a BNC cable between the BNC-to-banana adapters.
CX1101A/CX1206A as shown below.
a. Press the
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
b. Click User Calibration to display the Configuration > User Calibration
screen.
c. Set the ON/OFF switch to OFF for the channel connected to the
CX1101A/CX1206A.
In the example of
Figure
to OFF to disable the user calibration.
and ON as shown below.
a. Press the
key located in the Trigger control on the CX3300 front
Menu
panel. And set the CX3300 as follows.
In the following condition, Channel N is the channel connected to the
CX1101A/CX1206A. Where, N is the channel number 1, 2, 3, or 4.
Using the CX1101A/CX1102A/CX1103A
3-2, the ON/OFF switch of the Channel 1 must be set
Performing User Calibration
Figure
3-2.