If You Use The Cx1105A; Required Accessories - Keysight CX3300 Series User Manual

Device current waveform analyzer
Table of Contents

Advertisement

2. Press the
3. Press the
4. Set the following conditions on the dialog box (Configuration > Calibration
5. Click the close (x) button located in the upper right corner on the
6. Press the
7. Use the Vertical control knob to adjust the vertical scale. You will see square
Figure 1-5

If You Use the CX1105A

Required accessories

Keysight CX3300 User's Guide, Edition 4
(yellow),
1
channel connected to the Aux Out terminal and enable the channel.
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
Menu
Output).
Output State: ON
Source Shape: Square
The +5 V square pulse (1 kHz, 50% duty cycle) is applied from the Aux Out
terminal which has 50  output impedance.
Configuration dialog box to close the dialog box.
key to start waveform acquisition.
Run
pulse as shown in
Figure
Calibration Output monitored using the CX1216A Resistive Sensor Head
Test adapter, 1 ea., furnished with the CX1105A
Coaxial cable (BNC plug to BNC plug), 1 ea., furnished with the CX3300
(green),
(blue), or
2
3
1-5.
Setting Up the CX3300
Verifying Basic Operation
(red) key associated with the
4
25