Preparing the product for use

6.3.4 Display of current charge level using remote control (optional accessory)

Only when "DUAL" marking is not present (see page 12)
The current charge level is displayed in the status bar of the remote control when it is connected and switched on:
7 Preparing the product for use

7.1 Alignment

The following alignment guidelines contain descriptions for connecting the knee joint to a prosthetic socket. In
principle, the alignment of the prosthesis is independent of the type of connection for the knee joint. In case of a
connection to an osseointegrated, percutaneous implant system, a socket is not used during bench alignment in
the alignment apparatus. In this case, the central proximal point on the prosthetic socket corresponds to the
trochanter of the thigh bone (see illustration in the section "Bench alignment in the alignment apparatus" see
page 30).
Ensure that possible flexion or adduction of the transfemoral residual limb can be compensated to a permissible
extent by an adapter approved by the implant manufacturer in the course of static alignment optimisation. Safe
functioning of the knee joint is only guaranteed with biomechanically correct alignment.

7.1.1 Settings with the "X-Soft" adjustment software

7.1.1.1 Introduction

The "X-Soft" adjustment software makes it possible to optimise the product settings for a patient. The adjustment
software provides step-by-step guidance through the adjustment process. After the settings are configured, the
data for them can be saved and printed for documentation. If needed, these data can be loaded and fed into the
product.
You can consult the integrated online help of the adjustment software for further information.
INFORMATION
The 4X1 X-Soft adjustment software, version 1.10 or higher, is required for correct alignment. If X-Soft is on
hand in version 1.0 or higher, it can be updated.
Updating the X-Soft adjustment software
1) Click "Help > About" in the menu bar of the Data Station when you are connected to the Internet.
→ The window opens with the versions of the previously installed programs and the manufacturer's address.
2) Click the "Check for updates" button in this window.
→ A search for updates of previously installed software products and components is performed via the Inter­
net.
3) If updates are available, click "Download" in the column on the right in order to download and save the
update.
4) Extract the "ZIP file" and execute it.

7.1.1.2 Data transfer between the product and the PC

Product settings using the adjustment software can only be made via Bluetooth data transfer. For this purpose, a
Bluetooth wireless connection must be established between the product and the PC using the "60X5 BionicLink
28
1.
– Charge level of battery for currently connected prosthesis
– Prosthesis is being charged
2. 68% – Charge level of battery for currently connected prosthesis, as a
percentage
Genium X3 - Bionic Prosthetic System 3B5-X3/3B5-X3=ST,Ge­
Quantity
Battery charge level
0
1
2
3
4
5
nium X3 3B5-2/3B5-2=ST
0%-10%
10%-30%
30%-50%
50%-70%
70%-90%
> 90%