2.3.3 Stop
2.3.4 Play/Pause
2.3.5 Settings
2.3.6 USB Icon
2.3.7 Distance Readout
M-HPQA-001
•
The Stop Button will stop the motion of the Helipath Quick Action stand
at any time
•
Pressing the Stop button will cancel all automated movements including
Homing, returning to programmed Test Positions, and Helipath Test
operations. These actions will need to be restarted from the beginning
•
The Play/Pause Button controls starts the Helipath oscillating test
•
Pressing the Play button programs the current location as the top Test
Position 1 and begins the downward motion of a standard Helipath Test
•
The Play button will turn into a Pause symbol after the Helipath test is
initiated. Pressing the Pause button will pause the Helipath test motion while
remembering the Test Position 1 programming location
•
The Settings Button will bring the user to the Settings Menu
•
The Settings menu will allow users to adjust the Screen Brightness, Update Firmware,
access the registration barcode, and calibrate the joystick
•
The USB icon indicates that a USB-B cable is plugged in
•
A USB-B cable is used in integrated mode with compatible viscometers
•
The distance readout indicates the current location relative to home
•
The distance is indicated in millimeters
H E L I PAT H Q U I C K A C T I O N L A B O R AT O R Y S TA N D - O P E R AT I O N A L M A N U A L
14