User Guide
Cleaning and Maintenance
General
Performing cleaning and maintenance regularly reduces the chances of equipment failure.
CAUTION! Personnel performing any maintenance or cleaning must familiarize
themselves with the
procedures.
Cleaning
All of the device's surfaces should only be cleaned using a mild soap solution and a wrung
cloth.
CAUTION! Never use hard tools or abrasive materials when cleaning any part of the
device.
WARNING! Never use cleaning agents containing chlorinated solvents or acetic or
phosphoric acid. These constitute a health hazard and could damage the
instrument.
Replacing filter, measuring gas hose and needle
For every 10 hours of measurements a filter replacement warning will appear.
You now have two options:
1. Replace the filter
10 hours of measurements.
2. Select Postpone warning. This will postpone the warning for 1 hour. This continues until
you select OK - filter replaced.
NOTE! Since the actual filter replacement cannot be checked by the instrument,
selecting "OK - filter replaced" will reset the 10 hour measurement timer even if
filter has not been replaced.
P/N 300042-I
05/2019
EN
"Safety Instructions" on page 9
and select OK - filter replaced. This will reset the counter for another
®
Dansensor
before attempting any of these
Warning
08/17/06
Time to replace sample filter
OK - filter replaced
Postpone warning
CheckMate 3
13:46