Troubleshooting; Vacuum Cannot Be Established - Ametek mocon Dansensor LeakPointer H2O User Manual

Table of Contents

Advertisement

20
User Guide

4. Troubleshooting

Vacuum cannot be established

Does ejector start when pushing start
button?
With ejector started and lid open, can
you feel/hear vacuum in air-vent hole?
Are sealing without damages and
correct mounted.
When closing the lid, does sealing seal
properly against lid without air gap?
When closing the lid and starting ejector,
can you feel vacuum from balance hose
inlet?
Is hose between vacuum gauge and lid
mounted correct?
Call MOCON Europe A/S Support Dept.
P/N 390225-C
11/2021
YES
YES
YES
YES
YES
YES
EN
Dansensor
Check that air supply is connected.
NO
Check that no hoses are bent.
Check that hoses are connected
according to flow drawing - see
connections" on page
Check that flow adjustment is fully open
Check that no hoses are bent.
NO
Check that hoses are connected
according to flow drawing - see
connections" on page
Check that flow adjustment is fully open
Clean ejector and make sure that silencer
is not blocked.
Mount sealing correct - see
NO
of parts" on page 26
Replace sealing - see
details.
Adjust lid - see
NO
24
Remove balance hose
NO
Check that check valve in quick coupler
works. If not replace quick-coupler - see
page 32
for ordering details.
Mount hose according to flow drawing -
NO
see
"Internal connections" on page
LeakPointer H
O
®
2
"Internal
29.
"Internal
29.
"Replacement
page 32
for ordering
"Lid adjustment" on page
29.