APPENDIX B - FAULT DIAGNOSIS AND TROUBLESHOOTING

Error #
Description
0
RTC BATTERY VOLTAGE LOW
1
RTC RESET OCCURRED
2
SUPPLY VOLTAGE LOW
3
EEPROM INIT FAILED
4
RTC INIT FAILED
5
ACCEL INIT FAILED
6
STEPPER MOTOR DRV INIT FAILED
7
TORQUE SENSOR INIT FAILED
8
RTD INIT FAILED
9
INSTRUMENT RESET DUE TO HARD FAULT EXCEPTION
10
INSTRUMENT RESET DUE TO MEMORY FAULT EXCEPTION
11
INSTRUMENT RESET DUE TO BUS FAULT EXCEPTION
12
INSTRUMENT RESET DUE TO USAGE FAULT EXCEPTION
13
INSTRUMENT RESET DUE TO STACK OVER FLOW
14
FILESYSTEM WAS MISSING. REFORMAT FAILED
15
FILESYSTEM WAS MISSING. NEW INSTANCE CREATED
16
FILESYSTEM WAS MISSING. NEW INSTANCE CREATE FAILED
17
CHECKDISK REPORTED FAT CHAIN ERRORS
18
CHECKDISK REPORTED DIRECTORY ERRORS
19
CHECKDISK REPORTED LOST CLUSTER ERRORS
20
CHECKDISK REPORTED FILE SIZE ERRORS
21
TOUCH SCREEN INITIALIZATION FAILED
22
SYS TIME INITIALIZATION FAILED
23
SDRAM INITIALIZATION FAILED
24
RTC TIME INVALID
25
ADC INIT FAILED
26
LCD DRIVER INIT FAILED
27
MOTOR FAULT RESTART DEVICE
28
UPPER LIMIT SWITCH HIT UNEXPECTEDLY
M-HPQA-001
H E L I PAT H Q U I C K A C T I O N L A B O R AT O R Y S TA N D - O P E R AT I O N A L M A N U A L