Electrical Preparations And Connections - Ametek ta7000R User Manual

Trace analytical gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

instrument from air conditioners or extreme temperature fluctuations in
the room.
2.
lyzer. A uniform temperature environment for cylinders and regulators
is essential in order to keep constant pressure and flow rate. Always keep
gas containers inside the building where temperature fluctuations are at a
minimum.
3.
any high flow velocity draft directly onto the analyzer. Be sure not to
block any vent slots in the rear or top of the analyzer since limited air
circulation will cause overheating and poor performance.
4.
inch stainless steel tubing (maximum 0.030" ID). Coil some extra 1/16th
inch gas lines behind the analyzer. This allows flexibility when moving or
turning the instrument for service and routine maintenance.
5.
the rear panel for easy hook up of the gases, as well as ease of service
components. Allow approximately 30 inches service clearance at the rear
of the analyzer.

Electrical Preparations and Connections

1. Always verify that the voltage on the label located on the rear of the
instrument matches the supplied power before connecting. The electri-
cal requirements for the Reduction Gas Analyzer can be 100, 115, 220, or
240 VAC, and 50 or 60 Hz, depending on the voltage option ordered. AC
power must be free from cyclic RMS voltage variations, spikes or drops
greater than 1.5%. A minimum of 5 amperes current must be available.
2. Good grounding of the AC power circuit is required, verify proper
receptacle grounding before connecting the instrument.
3. It is recommended to connect the analyzer and low powered accesso-
ries to a dedicated electrical circuit. Do not plug the analyzer into a circuit
that is shared with other equipment with large intermittent current loads.
Line voltage spikes can have detrimental effects on the analytical capabili-
ties. Where there are noisy circuits, a power line conditioner should be
considered to even out spikes and other fluctuations.
4. Frequent power failures in the area where the analyzer is used are best
taken care of by installing an uninterruptible power supply (UPS). The
UPS will supply power instantly from batteries when the main power is
interrupted. Proper power will be supplied for minutes or hours depend-
Support gases for the instrument should be located near the ana-
Have adequate ventilation around the instrument area without
All connections to the instrument should be made with 1/16th
Situate the analyzer in such a way that there is always access to
User Manual | 25

Advertisement

Table of Contents
loading

Table of Contents