Spirometry setup

Select Setup Loops from the Spirometry menu
to access the Setup loops menu.
1. Spirometry Scaling: Set the scales of the
spirometry loop graph. The available
settings for the volume, Paw, and flow
graph axis are dependent on the set
patient type of adult or pediatric.
Auto: automatically adjusts the
volume, Paw, and flow axis of the loop.
Linked: links the adjustment of the
volume, Paw, and flow axis of the loop
graph together.
Indep. allows the axis of the loop
graph to be changed separately for
the volume, Paw, and flow axis.
2. Patient and Sensor: Select Adult or Pedi
depending on the sensor used.
3. Data Source: Select Patient to have
spirometry data sourced from the airway
module or Vent to have spirometry data
sourced from the ventilator.
4. Show MV or TV: Set to TV to show TVinsp
and TVexp on the spirometry split screen.
Set to MV to show MVinsp and MVexp on
the spirometry split screen.