Maintenance intervals
Component
Bulb for procedure
light/night light
See the "Overview"
chapter in the
associated instructions
for use.
Heating
elements/infrared
radiators
Bulb for phototherapy
light
Hose grommets
Repairs
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
Instructions for use Babytherm 8004/8010
Interval
Task
In case of fault
Exchange
Let the bulbs cool down for
at least 15 minutes.
After 2 years, see
Exchange in pairs
also Page 103.
After 1000
Exchange to preserve the
operating hours
specified therapeutic effect.
Only use listed bulbs (see
''Technical data'').
Always replace all bulbs.
In case of fault
Replace as soon as the
material becomes brittle
and/or sticky.
Service
Target group
Service personnel
Specialized service
personnel
Specialized service
personnel
User