Installation and Service

To Test the Module

Seat the cards and tighten the thumbscrews.
5
Starting with the bottom card, firmly seat the cards into the backplane
connector of the mainframe. Keep applying pressure to the center of the
card endplate while tightening the thumbscrews finger-tight. Repeat this for
all cards and filler panels starting at the bottom and moving to the top.
For correct air circulation, filler panels must be installed in all unused card
W A R N I N G
slots. Correct air circulation keeps the instrument from overheating. Keep
any extra filler panels for future use.
To Test the Module
If you require a test to initially accept the operation, perform the self-
tests later in this chapter.
If you need to deskew the channel-to-channel variations, go to
chapter 10, "Skew Adjust."
If you require a test to verify the specifications, start at the beginning
of chapter 3, "Testing Performance," in the optional Agilent
Technologies 16517A/18A Service Guide. The Service Guide is
available through your Agilent Technologies Sales Office.
If the module does not operate correctly, go to the beginning of
chapter 5, "Troubleshooting," in the optional Agilent Technologies
16517A/18A Service Guide. The Service Guide is available through
your Agilent Technologies Sales Office.