To Calibrate The Data Acquisition Module - Agilent Technologies 16517A User Reference

4-gsa/s timing and 1-gsa/s synchronous state logic analyzer
Table of Contents

Advertisement

Skew Adjust
This chapter gives you instructions for calibrating and adjusting the
16517A/18A module. Though the specifications are not tested during
these procedures, the module must be adjusted to ensure the
specifications will be met.
The adjustment to the high-speed logic analyzer consists of skew
adjustment to minimize channel-to-channel skew. The adjustment
factors are stored in nonvolatile RAM, and are loaded up at system
start up.
To periodically verify the performance of the 16517A/18A against the
module specifications, refer to "Testing Performance" in the optional
Agilent Technologies 16517A/18A Service Guide, available from
your Agilent Technologies Sales Office.

To Calibrate the Data Acquisition Module

The 16517A/18A does not require an operational accuracy calibration. To
test the module against the module specifications, refer to "Testing
Performance" in the optional Agilent Technologies 16517A/18A Service
Guide, available from your Agilent Technologies Sales Office.