Agilent Technologies 16517A User Reference page 172

4-gsa/s timing and 1-gsa/s synchronous state logic analyzer
Table of Contents

Advertisement

Edge Width: 16/32/48/64/80 channels.
Edge Counting Frequency: 444 MHz.
Edge Detection: Up to 1 GHz.
Greater than Duration (Timing only): 0 ns – 510 ns, accuracy is ±
2.25 ns.
Less than Duration (Timing only): 4 ns – 510 ns, accuracy is ±
2.25 ns.
Qualifier: A user-specified term that can be any state, no state, any
recognizer (patterns or edges), the timer, or the logical combination
(AND, OR, XOR) of the recognizers and timers.
Branching: Each sequence level has two branching qualifiers. When
satisfied, the analyzer will branch to the specified sequence level.
Maximum Occurrence Count: 16,777,216.
Maximum Sequencer Speed: 500 MHz.
State Sequence Levels: 4 plus trace point.
Timing Sequence Levels: 4 plus trace point.
Timer/Counter: There is one timer or counter per sequence level,
which is restarted upon entry into each level.
Timer/Counter Range:
Timing mode
0 s to 33 ms.
State mode
500 MHz to 1 GHz, (user clock period)×(2 to the 23rd).
below 500 MHz, (user clock period)×(2 to the 24th).
Timer Resolution:
Timing mode
2 ns.
above 500 MHz, 2 × user clock period.
State mode
below 500 MHz, user clock period.
Timer Accuracy: 0.005% of timer value.
Specifications and Characteristics
Characteristics
12–7

Advertisement

Table of Contents
loading

This manual is also suitable for:

16518a