Statistics Markers - Agilent Technologies 16517A User Reference

4-gsa/s timing and 1-gsa/s synchronous state logic analyzer
Table of Contents

Advertisement

Statistics Markers

After patterns are assigned to the X and O markers, statistical
information becomes available when markers are set to Statistics. The
logic analyzer displays the following information:
Number of valid runs (runs where Pattern markers were able to be
placed on specified patterns).
Minimum time between the X and O Pattern markers.
Maximum time between the X and O Pattern markers.
Average time between the X and O Pattern markers.
The markers are color coded for easy recognition in the data listing.
The X marker is represented by a green line and the O marker is
represented by a yellow line.
Statistics are based on the time between the X and O. Both markers
must be found before valid statistical information is displayed.
In repetitive run mode, the display is updated each time a valid run
occurs until you select Stop.
In single run mode, each time you select Run an additional valid run
will be added to the data and the statistics will be updated. This
process continues unless you change the placement of the X and O
Pattern markers between runs.
9–14