The Trigger Menu
Assigning Resource Term Names and Values
Timer Term
Timer options are selected from within sequence levels. The timer term is
evaluated as either true or false. A true timer term lets the sequence
evaluation continue within the current statement. Because the timer works
exclusively for each level it is assigned to, it will not affect any other level or
the trigger specification from a global respect. However, each sequence level
can have its own timer with different values assigned. The timer term
options are used in the following ways:
"Timer expired" The timer term becomes true when its assigned time
expires.
"Timer not expired" The timer term is true while the assigned time
has not expired. Once expired, the timer term becomes false. This is
like negating the timer.
Combination of Terms
Combination terms are configured and selected from within sequence levels.
All user-defined resource terms can be combined to create complex qualifiers
that occupy a single assignment field space.
When the term assignment field is selected, and "Combination" is selected
from the pop-up selection list, a logical assignment menu appears. Use this
menu to select pre-defined resource terms as inputs to a chain of logical
operators.