Error Messages
Skew Adjust and Performance Verification Messages
Warning Messages
Skew adjustment halted, recalling stored values. Calibration was
halted before it completed.
Skew adjustment failed, recalling stored values. This message
appears when any calibration fails.
Skew adjustment failed. Press "Proceed" to retry or "Exit" to quit.
This message appears when any calibration fails, but only before the user
retries or quits.
16517/18 Skew values do not match HW configuration. The
hardware configuration corresponding to the calibration factors stored do
not match the current configuration in the frame.
Using Default 16517/18 Skew Values. Please See "Skew Adjust"
Menu.
Since no valid calibration values were found in flash ROM, or those read
do not match the current hardware, the default values are being loaded.
The board skews should be re-adjusted.
Advisory Messages
Previous configuration: A:16518 B:16518 C:16517 D:16518
E:16518. This message describes the configuration that produced the stored
calibration factors.
Please adjust the skew or re-configure the module. This message
appears after CAL_HW_MISMATCH, CAL_CONFIG and DFT_CAL_MSG
are detected. Adjust the skew, or reconfigure the module.
The clock is not connected properly. The module was unable to
measure a signal on the clock as part of the calibration process.
Data channels (........) are not connected properly. The module was
unable to measure a signal on any of the data channels as part of the
calibration process.
11–9