Environmental Requirements
5.2
Environmental Requirements
5.2.1
Transport/Storage and Operating Conditions
Local Restrictions
• Store and operate systems in locations that are protected against moisture, wind, sunlight,
dust, salinity and sulfur.
• Do not store systems in the close vicinity of chemical products or gas.
• Do not expose systems to direct UV light.
Altitude Considerations
• Unless stated otherwise, the system is rated for use at an altitude < 3000 meters.
• Transport or store the system at an altitude < 6000 meters.
Adaption Time
Adaption time for use after extreme storage conditions is a minimum of one hour.
Monitor and Camera Carts
The following environmental requirements apply for the Curve system:
Temperature
Humidity
Pressure
56
Transport/Storage Conditions
-10°C (14°F) to 50°C (122°F)
10% to 90% non-condensing
500 hPa to 1060 hPa
Technical User Guide Rev. 1.2 Curve Dual Display, Curve Single Display Ver. 1.0
Operating Conditions
10°C (50°F) to 35°C (95°F)
30% to 75% non condensing
700 hPa to 1060 hPa