Using the Removable Ion Source
To Change the Ion Source
The steps necessary for changing the ion source are provided here and also
interactively in the MassHunter Workstation software. The process is mostly
automated, with the exception of the use of the RIS probe extraction tool to
remove and install the ion source. Purging the chamber, cooling the source,
and pumping down are all performed with the use of the Workstation
software.
Procedure
1
2
Figure 19
GC/Q-TOF Tune dialog box
78
From MassHunter Data Acquisition software control view, click the MS
Tune icon to display the GC/Q-TOF Tune dialog box.
Click the Removable Ion Source tab in the GC/Q-TOF Tune screen.
7200 Accurate-Mass QTOF Operation Manual
(Figure
19)