Two-Point Calibration Of The Standard Signal General - ProMinent DULCOMETER D1Cb Series Instructions For Assembly And Use

Hide thumbs Also See for DULCOMETER D1Cb Series:
Table of Contents

Advertisement

Measured Variables and Operating Menus for the Standard Signal General
Calibration in the reduced operating menu
1.
Select the calibration menu. Then press
2.
Enter the actual value suitable for the respective measured variable determined using the
measuring method with the keys
the entry with
3.
Then press
menu " on page 91
Calibration is completed.
ð
4.
Then press
The display now shows the values determined for the zero point and gradient.
ð
8.8.2

Two-Point Calibration of the Standard Signal General

calibration
zero p.:
4.00 mA
slope
0.16 mA
Permanent display 1
Fig. 45: Zero point calibration menu / shown with flow as the measured variable and the measuring
range 0...100 m
/h
3
92
,
. Should an error be displayed, see
calibration
mea. val
1
take over value
2m
3
/h 5.04 mA
calibration
mea. val
2
take over value
80m
3
/h 13.17 mA
calibration
zero p.:
3.98 mA
slope
0.11 mA
and
into the DULCOMETER
Ä Table "Error messages in the calibration
calibration
3
value1 :
2m
/h
value2 :
calibration
value1 :
2m
3
/h
80m
3
/h
value2 :
®
D1Cb. Confirm
D1C2-mA-030-GB