Calibrating the ORP meas‐
9.2
ured value
Selection of the calibration
9.2.1
process for ORP
Calibration process selection
To calibrate the controller there are two
available calibration processes:
Single point calibration (with buffer
n
solution)
Data input
n
1.
Continuous display ➨
CAL ORP
Last calibration
0.0 mV
Offset
Select calib. proc.
PA installed
continue with <CAL>
Fig. 41: Calibration menu [ORP]
ð The calibration menu is dis‐
played.
2.
Using the arrow keys select the
desired menu item. Press
ð The menu for selecting the cali‐
bration process appears.
3.
Using the arrow keys select the
desired menu entry
[Select calib. proc.] and press the
key
ð The entry window appears.
4.
Using the arrow keys select the cal‐
ibration process and press the
key
5.
Then press
Single point calibration
13:26:11
11/11/2011
Yes
ð You can now start the selected
calibration process.
Single point calibration ORP
9.2.2
sensor (CAL)
Correct sensor operation
Correct measuring and metering
–
is only possible if the sensor is
working perfectly
Observe the sensor operating
–
instructions
For calibration the sensor must
–
be removed and refitted in the in-
line probe housing. To do this,
observe the operating instructions
of your in-line probe housing
ORP sensor calibration
A1027
The ORP sensor cannot be cali‐
brated. Only a deviation [OFFSET]
with a value in the range ± 40 mV can
be set and thereby compensated. If
the ORP sensor deviates by more
than ± 40 mV from the reference
value, then it must be checked in
accordance with the requirements of
the sensor operating instructions.
Calibrating the controller
73