Measured Variables and Operating Menus for the Standard Signal General
general setting
information
Fig. 40: Reduced / complete changeover

8.4 Setting the Reading

WARNING!
Incorrect metering due to incorrect measured variable
Possible consequence: Fatal or serious injuries
The measured variable/measuring range of the sensor/measuring device is essential for the
measured variable/measuring range!
If the measured variable/measuring range is changed, the sensor must be recalibrated
If the measured variable/measuring range is changed, the set points and limit values will be
changed to the associated starting values!
If the measured variable/measuring range is changed, the settings in all of the menus must
be checked
measured value
setting ?
Fig. 41: Reading
84
Identcode:
D1CB xxxxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxxx
Standard signal input mA
range adjustment
measured value
4 m A
=
0.00 ppm
alarm relay
operating menu
Calib timer
wash timer
range adjustment
measured value
20 mA =
2 .00 ppm
operating menu
_
reduced
_
english
5000
access c.:
A0088_GB
A0019_GB