Measured Variables and Operating Menus for Amperometric Sensors
6.7 Setting the Reading
WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
–
The measuring range of the sensor is essential for the measuring range!
–
If the assignment of the measuring range is modified, the settings must be checked in all
menus
–
If the assignment of the measuring range is changed, the sensor must be recalibrated
–
The relevant information can be found in the operating manual for the sensor/measuring
equipment
measured value
setting ?
Fig. 30: Reading
This setting is solely used to adapt the DULCOMETER
providers. Sensors provided by third party providers may have measuring ranges, which deviate from
the DULCOMETER
®
In order to adapt a ProMinent sensor to the DULCOMETER
in "General Settings", see
Range " on page 37
6.8 Correction Value
Only necessary when using the
DULCOTEST
CDP sensor for chlorine
®
dioxide CIO
.
2
60
Standard signal input mA
range adjustment
measured value
4 m A
=
0.00 ppm
D1Cb standard specification.
Ä Chapter 4.1.2 "Selection of the Measured Variable and Measuring
range adjustment
measured value
20 mA =
2 .00 ppm
®
D1Cb to the sensors provided by third party
®
D1Cb only use the menu to be found
only with correction value
correcting value
correction value
temperature
setting ?
automatic
Fig. 31: Correction Value
A0019_GB
A0020_GB