Setting The Reading; Correction Value - ProMinent DULCOMETER D1Cb Series Instructions For Assembly And Use

Hide thumbs Also See for DULCOMETER D1Cb Series:
Table of Contents

Advertisement

Measured Variables and Operating Menus for Amperometric Sensors

6.7 Setting the Reading

WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
The measuring range of the sensor is essential for the measuring range!
If the assignment of the measuring range is modified, the settings must be checked in all
menus
If the assignment of the measuring range is changed, the sensor must be recalibrated
The relevant information can be found in the operating manual for the sensor/measuring
equipment
measured value
setting ?
Fig. 30: Reading
This setting is solely used to adapt the DULCOMETER
providers. Sensors provided by third party providers may have measuring ranges, which deviate from
the DULCOMETER
®
In order to adapt a ProMinent sensor to the DULCOMETER
in "General Settings", see
Range " on page 37

6.8 Correction Value

Only necessary when using the
DULCOTEST
CDP sensor for chlorine
®
dioxide CIO
.
2
60
Standard signal input mA
range adjustment
measured value
4 m A
=
0.00 ppm
D1Cb standard specification.
Ä Chapter 4.1.2 "Selection of the Measured Variable and Measuring
range adjustment
measured value
20 mA =
2 .00 ppm
®
D1Cb to the sensors provided by third party
®
D1Cb only use the menu to be found
only with correction value
correcting value
correction value
temperature
setting ?
automatic
Fig. 31: Correction Value
A0019_GB
A0020_GB