Measured Variables and Operating Menus for the Standard Signal General
8.2 Changing the Measured Variable
WARNING!
Incorrect metering due to incorrect measured variable
Possible consequence: Fatal or serious injuries
–
The measured variable/measuring range of the sensor is essential for the measured vari‐
able/measuring range!
–
If the measured variable/measuring range is changed, the sensor must be recalibrated
–
If the measured variable/measuring range is changed, the set points and limit values will be
changed to the associated starting values!
–
If the measured variable/measuring range is changed, the settings in all of the menus must
be checked
general setting
informationen
Fig. 39: Changing "Standard signal general" measured variable
For possible measured variables see
8.3 Reduced / Complete Operating Menu
The DULCOMETER
®
(reduced / complete). All of the parameters of the control unit are preset and can be changed in the
complete operating menu.
The controller is delivered with a reduced operating menu. If adjustments are necessary, the param‐
eters can be changed by switching to the complete operating menu.
Identcode:
D1CBxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxx
Ä Table on page 85 .
D1Cb allows settings to be made in two different comprehensive menus
Calib. timer
wash timer
measure change
add. functions
PAA
H2O2 (PER)
mA general
pH over mA
effective range
0...105 %
A0232_GB
83