9.7 General Settings

general setting
information
Fig. 61: General Settings
The functions described below can be selected in this menu.
9.7.1

Setting the Measured Variable/Measuring Range

WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or very serious injuries
If the assignment of the measuring range is modified, the settings must be checked in all
menus
If the assignment of the measuring range is changed, the sensor or the measuring device
must be recalibrated
The measuring range of the sensor or the measuring device is essential for the measuring
range!
general setting
informationen
Fig. 62: Measuring range
Operating Menus Independent of Measured Variables
identcode
D1CBxxxxxxxxxx
Vxxxxxxxxxxx
Srnr: xxxxxxxxxx
Identcode:
D1CBxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxx
alarm relay
pause
operating menu
Calib timer
calib timer
wash timer
measure change
temperature °C-°F
operating menu
reduced
-
-
English
access c.:
5000
A0016_GB
ORP
---------------------
Chlor
Chlordioxid
effective range
0...2.00ppm
A0032_GB
109