Two-Point Calibration Of The Standard Signal General - ProMinent DULCOMETER D1Cb Assembly And Operating Instructions Manual

Hide thumbs Also See for DULCOMETER D1Cb:
Table of Contents

Advertisement

Measured Variables and Operating Menus for the Standard Signal General
Calibration in the restricted operating menu
1.
Select the calibration menu. Then press
2.
Enter the actual value suitable for the respective measured variable determined
using the measuring method with the keys
D1Cb / D1Cc. Confirm entry with the key
3.
Continue with
bration menu' Table on page 110
ð Calibration is completed.
4.
Continue with
ð The display now shows the values determined for the zero point and slope.
9.8.2

Two-Point Calibration of the Standard Signal General

calibration
zero p.:
4.00 mA
slope
0.16 mA
Permanent display 1
Fig. 54: Two point calibration menu / shown with flow as the measured variable and the
measuring range 0...100 m
. Should an error be displayed, see
calibration
mea. val
1
take over value
2m
3
/h 5.04 mA
calibration
mea. val
2
take over value
80m
3
/h 13.17 mA
calibration
zero p.:
3.98 mA
slope
0.11 mA
/h
3
,
and
in the DULCOMETER
Ä 'Error messages in the cali‐
calibration
3
value1 :
2m
/h
value2 :
calibration
value1 :
2m
3
/h
value2 :
80m
3
/h
®
D1C2-mA-030-GB
111