4.1.2

Selection of the Measured Variable and Measuring Range

WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
The measuring range of the sensor is essential for the measuring range!
If the assignment of the measuring range is modified, the settings must be checked in all
menus
If the assignment of the measuring range is changed, the sensor must be recalibrated
With devices, which have not been preconfigured to the customer's specific requirement, the measured
variable required has to be selected in the complete operating menu "General Settings / Change
Measured Variable". The DULCOMETER
to the measured variable selected. The relevant labels are enclosed with the
DULCOMETER
D1Cb .
®
general setting
informationen
Fig. 15: Selection of measured variable and measuring range
The measuring range required has to be selected and set in the full operating menu "General
Settings / Change Measured Variable/", see
uring Range " on page 109 .
®
D1Cb then has to be labelled with the label corresponding
Identcode:
D1CBxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxx
Ä Chapter 9.7.1 "Setting the Measured Variable/Meas‐
calib timer
wash timer
measure change
temperature °C-°F
Commissioning
ORP
---------------------
Chlor
Chlordioxid
effective range
0...2.00ppm
A0032_GB
37