Calibration Of The Zero Point Of The Standard Signal General - ProMinent DULCOMETER D1Cb Series Instructions For Assembly And Use

Hide thumbs Also See for DULCOMETER D1Cb Series:
Table of Contents

Advertisement

Measured Variables and Operating Menus for the Standard Signal General
During calibration, the DULCOMETER
this is if a basic load or a manual actuating variable has been set. This remains active. The mA standard
signal outputs are frozen. The reading frozen at the start of calibration is suggested as a value. This
value can be set using the arrow keys. Calibration is only possible if the value is ≥ 2 % of the measuring
range of the sensor or the measuring device.
Error messages in the calibration menu
Error message
Reading gap too small
Refer to the operating manual of the respective sensor for all error messages and use for trouble‐
shooting.
* of the measuring range and measured value, see
8.8.1

Calibration of the Zero Point of the Standard Signal General

calibration
zero p.:
4.00 mA
slope
16.00 mA
Permanent display 1
Fig. 44: Zero point calibration menu / shown with flow as the measured variable and the measuring
range 0...100 m
/h
3
The gradient in the calibration menu is given as an mA value. This value should be regarded as
the mA/selected measuring unit.
D1Cb sets the actuating outputs to "0". The exception to
®
Condition
∆ Value < 5.0 %*
∆ Value < 1.00 mA*
calibration
mea. val
00003 m
3
/h
Effect
Reading is rejected
Repeat calibration of the meas‐
uring point!
Ä Table on page 85
calibration
zero p.:
3.98 mA
slope
16.02 mA
D1C2-mA-029-GB
91