Measured Variables and Operating Menus for the Standard Signal General
9.3
Reduced / Complete Operating Menu
The DULCOMETER
hensive menus (reduced / complete). All of the parameters of the controller are preset
and can be changed in the complete operating menu.
The controller is delivered with a restricted operating menu. If adjustments are neces‐
sary, the parameters can be changed by switching to the complete operating menu.
general setting
information
Fig. 49: Reduced / complete changeover
9.4
Setting the Reading
WARNING!
Incorrect metering due to incorrect measured variable
Possible consequence: Fatal or serious injuries
–
The measured variable/measuring range of the sensor/measuring device is
essential for the measured variable/measuring range!
–
If the measured variable/measuring range is changed, the sensor must be recali‐
brated
–
If the measured variable/measuring range is changed, the set points and limit
values will be changed to the associated starting values!
–
If the measured variable/measuring range is changed, the settings in all of the
menus must be checked
measured value
setting ?
Fig. 50: Reading
104
®
D1Cb / D1Cc allows settings to be made in two different compre‐
Identcode:
D1CB xxxxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxxx
Standard signal input mA
range adjustment
measured value
4 m A
=
0.00 ppm
alarm relay
operating menu
Calib timer
wash timer
range adjustment
measured value
20 mA =
2 .00 ppm
operating menu
_
reduced
_
english
5000
access c.:
A0088_GB
A0019_GB