1. Place the projector on a clean horizontal surface.
2. On the IR remote or keypad, press Menu > Configuration > Lens Setting.
3. Select Lens Calibration.
4. Press Enter.

ADC calibration

If the color appears abnormal after upgrading the scalar firmware, perform an ADC calibration.
Warning! Failure to comply with the following could result in death or serious injury.
• Do not look into the projector lens when the laser is on. The extremely high brightness can cause
permanent eye damage.

Performing a PC calibration

To perform the PC calibration, complete the steps below.
ID
Part name
1
Test equipment
2
Test signal
3
Test pattern
4
Inspection item
1. On the IR remote keypad, press Menu > Size & Position > Size Presets > Full Size.
2. Enter Factory mode.
See
Entering Factory mode
3. Select ADC Calibration > Help.
GS Series 700-850 Service Manual
020-001054-02 Rev. 1 (04-2017)
Description
The lens shift is moving smoothly.
Description
Video generator (output level: 700 mv)
Timing: 800 x 600 @60Hz
16 Gray
Check if the ADC calibration status has changed from Doing to Done.
on page 13.