Performing The Dac Calibration - Christie Cinema 4K-RGB Service Manual

Hide thumbs Also See for Cinema 4K-RGB:
Table of Contents

Advertisement

7. Install the new color sensor board harness (P/N: 001-113506-02) starting near the
housekeeping board, routing through all routing clips and finishing at the lowest P-clip.
8. Gently route the color sensor board harness under the IOS and pull it into position at the
underside of the projector.
9. Connect the harness to the color sensor board.
10. Re-install the color sensor board. modules, and covers previously removed.

Performing the DAC calibration

Calibrate the black levels for the new color sensor board.
1. Prepare the projector for DAC calibration:
a) Turn off the laser light.
b) Close the shutter.
c) Install the lens cap.
The lens cap minimizes any stray light entering through the lens and maximizes darkness
around the sensor during the calibration.
2. Enable direct communication to photon controller.
Cinema 4K-RGB Service Guide–CP4315-RGB, CP4320-RGB, CP4325-RGB, CP4330-RGB
020-102713-06 Rev. 1 (08-2022)
Copyright
©
2022 Christie Digital Systems USA, Inc. All rights reserved.
If printed, please recycle this document after use.
Printed circuit boards and sensors
99