5. To confirm the deletion, select Delete.
LiteLOC
sensor-to-screen calibration
The LiteLOC sensor-to-screen calibration is performed in the factory under specific setup conditions
which include the type of lens, screen, and the spectroradiometer used for measuring screen color and
brightness.
Your projector's setup conditions may not match the factory setup which may result in an discrepancy
in color accuracy. Upon first installation, check the accuracy of the color point and if unsatisfactory,
perform this calibration. Doing this calibration captures all the variables of your setup and achieves the
best results. By doing this calibration, you will not overwrite the factory calibration. Once satisfied,
some scenarios may exist in which you may need to re-do the calibration in the future.
Re-doing this calibration is required if the following is replaced:
Color sensor board
Christie recommends re-doing this calibration if any of the following are replaced:
Light engine
Lens
LOS (laser optical subsystem)
Screen
Port window
Any other optical components in the optical path between the LOS and the screen
After replacing these components, check the accuracy of the color point and if unsatisfactory, re-do
this calibration.
LiteLOC
sensor-to-screen calibration: Workflow
Identify the recommended order for performing the sensor-to-screen calibration.
LiteLOC
sensor-to-screen calibration requires the following:
Spectroradiometer
Service user is the minimum level user access
Create a new calibration file (on page 35).
Recalibrate the calibration file (on page 37), if required.
Edit the existing calibration file parameters (on page 38) on the Sensor-to-screen menu, if
required.
Verify the calibration (on page 38), if required.
Add the calibration file to the Channel Setup (on page 39).

Creating a new calibration file

The Sensor-to-screen Calibration wizard guides you through the steps required to create a new
calibration file.
1. Set up a spectroradiometer to obtain screen measurements later in the process.
Cinema 2K-RGB Installation and Setup Guide–CP2415-RGB and CP2420-RGB
020-103259-01 Rev. 1 (10-2022)
Copyright
©
2022 Christie Digital Systems USA, Inc. All rights reserved.
If printed, please recycle this document after use.
Managing the light source
35