Removal and Replacement Procedures
Post-Repair Procedures
OF Ingot Utilities
Overview of the OF Ingot Utilities
The OF Ingot Utilities is a software used to program back all the necessary
information back into the P937xA USB VNA. It should be performed whenever
replacing a new chassis or module replacement.
Procedure
Step 1. Once the P937XA has been re-assembled, power up the USB VNA.
Step 2. Connect the chassis to the System Controller (PC) via USB cable.
a. Consult the user guide (P9370-90001) for the minimum requirements of the
Step 3. Download and install the "OF Ingot Utilities".
You will need to contact the Keysight factory support node to request the
OF Ingot Utilities software.
Step 4. Run the OF Ingot Utilities.
Step 5. When OF Ingot Utilities opens the main menu with look like the image below.
a. Make sure you see the host id in the Connected Chassis window.
74
system controller.
Keysight P937xA PXIe Service Guide