Removal and Replacement Procedures

Information in This Chapter

Keysight P937xA PXIe Service Guide
5
Removal and Replacement Procedures
STOP!!! Please read all content before proceeding with this repair
procedure. This repair must be done at a service center! If you think you
may not be qualified to do the work, or need advice, contact Keysight.
Refer to
Information in This Chapter
This chapter contains procedures for removing and replacing the major
assemblies of your Keysight USB Streamline network analyzer.
Section Title
Electrostatic Discharge (ESD)
Protection
Connector Replacement Procedure
Front Panel Replacement Procedure
Chassis Replacement Procedure
Module Replacement Procedure
Post-Repair Procedures
The hardware does not support "hot-swap" (changing network analyzer
modules while power is applied to the chassis) capabilities. Before
installing or removing a module to/from the chassis, power-off the chassis
to prevent damage to the module.
"Keysight Support, Services, and Assistance" on page
Summary of Content
Information pertaining to ESD protection.
A procedure for replacing the Streamline connectors.
Procedure for replacing the front panel.
Procedure for replacing the USB-VNA chassis.
Procedure for replacing the module inside the
USB-VNA.
A table for the proper tests, verifications and
adjustments to perform on your analyzer after repair.
61
22.
Start Page
page 62
page 63
page 65
page 68
page 71
page 73